Reconstruction of 3D knee MRI using deep learning and compressed sensing: a validation study on healthy volunteers

Abstract

Background

To investigate the potential of combining compressed sensing (CS) and artificial intelligence (AI), in particular deep learning (DL), for accelerating three-dimensional (3D) magnetic resonance imaging (MRI) sequences of the knee.

Methods

Twenty healthy volunteers were examined using a 3-T scanner with a fat-saturated 3D proton density sequence with four different acceleration levels (10, 13, 15, and 17). All sequences were accelerated with CS and reconstructed using the conventional and a new DL-based algorithm (CS-AI). Subjective image quality was evaluated by two blinded readers using seven criteria on a 5-point-Likert-scale (overall impression, artifacts, delineation of the anterior cruciate ligament, posterior cruciate ligament, menisci, cartilage, and bone). Using mixed models, all CS-AI sequences were compared to the clinical standard (sense sequence with an acceleration factor of 2) and CS sequences with the same acceleration factor.

Results

3D sequences reconstructed with CS-AI achieved significantly better values for subjective image quality compared to sequences reconstructed with CS with the same acceleration factor (p ≤ 0.001). The images reconstructed with CS-AI showed that tenfold acceleration may be feasible without significant loss of quality when compared to the reference sequence (p ≥ 0.999).

Conclusions

For 3-T 3D-MRI of the knee, a DL-based algorithm allowed for additional acceleration of acquisition times compared to the conventional approach. This study, however, is limited by its small sample size and inclusion of only healthy volunteers, indicating the need for further research with a more diverse and larger sample.

Trial registration

DRKS00024156.

Relevance statement

Using a DL-based algorithm, 54% faster image acquisition (178 s versus 384 s) for 3D-sequences may be possible for 3-T MRI of the knee.

Key points

• Combination of compressed sensing and DL improved image quality and allows for significant acceleration of 3D knee MRI.

• DL-based algorithm achieved better subjective image quality than conventional compressed sensing.

• For 3D knee MRI at 3 T, 54% faster image acquisition may be possible.

Graphical Abstract

Background

The knee is the second most common site of musculoskeletal pain [1]. Common knee injuries, such as anterior cruciate ligament (ACL) injuries, are highly prevalent in the general population [2, 3], leading to decreased productivity and diminished quality of life [4], while also imposing a substantial socioeconomic burden [5]. In this context, magnetic resonance imaging (MRI), with its ability to provide detailed tissue contrast, plays a critical role in the accurate diagnosis and evaluation of knee injuries [6]. However, the growing demand for MRI [7] has led to a bottleneck in terms of patient throughput, limiting the number of patients who can undergo timely examinations. To expedite image acquisition without compromising image quality, several innovative techniques have emerged. These include reduced scanning protocols [8], parallel imaging (e.g., GRAPPA) [9], and compressed sensing (CS) [10]. Numerous studies have demonstrated the efficacy of CS in reducing scan times across various anatomical regions [11,12,13,14] while maintaining diagnostic image quality. However, conventional CS approaches face limitations in achieving higher acceleration factors due to the introduction of artifacts such as aliasing and blurring. Recent advancements in artificial intelligence (AI), especially deep learning (DL) have opened avenues for addressing these limitations by integrating it with conventional CS [15].

In the context of knee MRI, Pezotti et al. [15] introduced Adaptive-CS-Net as part of the 2019 fastMRI challenge which enabled an 8 × reduction in acquisition times. Adaptive-CS-Net is an AI-driven convolutional neural network that enhances MRI image reconstruction by integrating parallel imaging, CS, and DL, using iterative learning-based techniques to process undersampled k-space data, thereby improving image authenticity. Adaptive-CS-Net has since been further developed to accommodate a broader range of acceleration factors and anatomical regions, including the knee. In an initial study, Iuga et al. demonstrated a 64% reduction in scan time for two-dimensional knee MRI [16]. However, so far, there have been no studies testing the clinical feasibility and limitations of Adaptive-CS-Net to reconstruct three-dimensional (3D) MRI images of the knee. The implementation of this algorithm could help facilitate the routine clinical application of 3D-sequences, enabling enhanced visualization of anatomical structures not parallel to standard imaging planes, thereby improving diagnostic accuracy and efficiency. Thus, the aim of the current study was to compare the subjective image quality of MRI reconstructions using conventional CS and reconstructions using CS together with artificial intelligence (CS-AI) at different acceleration factors for 3D diagnostic MRI of the knee. The purpose was to determine whether CS-AI is a feasible approach to further decrease acquisition times of knee MRI while maintaining diagnostic image quality.

Methods

Study population

This prospective single center study was carried out in accordance with the ethical standards in the 1964 Declaration of Helsinki and its later amendments and was approved by our institutional review board. The study was registered in the national Clinical Trials Register (DRKS00024156). Recruitment of volunteers and acquisition of imaging data were carried out in March 2021. Written informed consent was obtained from all participants included in the study. Exclusion criteria were pregnancy, age below 18, implanted MRI conditional or unsafe devices, previous surgery or known pathologies of the knee, and knee related pain in the last 6 months.

MRI acquisition and reconstruction

A whole-body 3.0-T MRI system (Philips Ingenia 3.0 T, Philips, Hamburg, Germany) with a dedicated knee coil (transmit/receive, 16 channels) was used for image acquisition. All volunteers were placed supine, feet-first on the table. For all sequences, the field-of-view covered the entire knee joint. The protocol included a fat saturated 3D proton density sequence with four different acceleration levels (10, 13, 15, and 17) as well as a sense sequence with an acceleration factor of 2. These specific levels were chosen to represent a broad spectrum of acceleration, from moderate to extremely high, allowing us to assess the efficacy of the CS-AI algorithm in maintaining image quality under varying degrees of data undersampling, corresponding to scan time reductions of 54%, 64%, 68%, and 73%, respectively. The selected acceleration levels were compared with the conventional sense sequence at an acceleration factor of 2, which is the standard 3D sequence used in our institution’s daily routine. Except for the acceleration factors, all other parameters were kept identical between the acquired sequences. Table 1 summarizes the sequence parameters used in this study, showing the scan time for each sequence and the potential reduction of scan time.

Table 1 Acquisition parameters for the different three-dimensional sequences and results for changes in the scan time

The sets of undersampled k-space data from the different acceleration levels were reconstructed into visually perceivable images using two methods: (1) a conventional approach with CS; and (2) a novel AI-driven prototype (CS-AI). The CS-AI-prototype is based on the “Adaptive-CS-Net” convolutional neural network, which employs an iterative, learning-based reconstruction scheme to process undersampled k-space data by exploiting additional information such as coil sensitivity and location of the image background, resulting in a combined algorithm of parallel imaging, CS and DL. Unlike the traditional wavelet transformation, the neural network performs sparsifying and consistency checks on the raw k-space data within each block, resulting in maximized image authenticity [17]. The algorithm presented by Pezzotti et al. [15] was extended by using training data of about 740,000 images with various anatomies, contrasts and field strengths (1.5 and 3 T). Both the acquisition and reconstruction algorithms of CS and CS-AI were provided by the manufacturer.

Subjective image analysis

All scans were exported as Digital Imaging and Communications in Medicine − DICOM files to the clinical Picture Archiving and Communication System − PACS. Two board certified radiologists reviewed all images (R1, 8 years of experience; R2, 6 years of experience). For the subjective reading, the images were presented in random order and both readers were blinded to the scan sequence and reconstruction. All blinded images of a subject were available at once for both readers. Readers were free to choose window width and level settings and the review was performed over a period of 6 weeks.

Using a 5-point Likert scale, each reader independently evaluated delineation of the following anatomical structures for all sequences and reconstructions: anterior cruciate ligament, posterior cruciate ligament, menisci, cartilage and bone. Overall image impression and visible artifacts were rated additionally on a 5-point Likert scale, resulting in a total of seven subjective ratings for each of the nine images (CS 10/CS-AI 10, CS 13/CS-AI 13, CS 15/CS-AI 15, CS 17/CS-AI 17, and standard CS) reconstructed for every patient. Table 2 shows an overview of the used scale.

Table 2 Ratings for the anatomical structures, diagnostic certainty/overall image impression and artifacts used by the two readers for the evaluation of all sequences

Statistical analysis

GraphPad Prism version 9.0.1 for Mac OS X (GraphPad Software, Boston, United States) was used for all statistical analyses. For the subjective image analysis, the values from both readers for each sequence were averaged. To assess the interrater agreement between both readers, Krippendorff’s α was calculated, with a Krippendorff’s α ≥ 0.80 indicating high agreement, 0.67–0.79 indicating moderate agreement, and < 0.67 indicating poor agreement [18]. After assessing the normal distribution of the data using both graphical (e.g., Q-Q plots and histograms) and statistical methods (D’Agostino and Pearson test [19]), robust mixed models fitted with the restricted maximum likelihood method − REML were employed to analyze the effect of acceleration level (10, 13, 15, and 17) and reconstruction method (CS versus CS-AI) on indicators of subjective image quality (overall impression, artifacts, and delineation of the anterior cruciate ligament, posterior cruciate ligament, menisci, cartilage and bone). As post hoc tests, Sidak’s multiple comparisons test was used to compare the different reconstruction methods (CS versus CS-AI) at the different acceleration levels. Additionally, Dunnett’s multiple comparisons test was used to compare all sequences to the reference sequence (standard sense with an acceleration level of 2). All post hoc tests were corrected for multiple comparisons. Data are reported as the mean ± standard deviation. A p-value below 0.05 was considered statistically significant. A priori sample size calculation was performed using G*power 3.1.9.7 based on previous results for acceleration techniques in knee imaging [12]. A minimum number of 19 volunteers were needed to detect a difference of 0.2 points on the Likert scale with 0.3 standard deviation, with an alpha error of 0.05 and a power of 0.80.

Results

Study population

Twenty young, healthy volunteers (12 males and 8 females; mean age 30.5 ± 2.1 years, range 24−43 years; mean weight 74 ± 12 kg, range: 55−100 kg; mean height 176 ± 11 cm, range 155−90 cm) were included.

Image analysis

Scan time decreased with increasing acceleration factors; for instance, from 384 s for the standard sequence to 178 s for an acceleration factor of 10 and down to 105 s for an acceleration factor of 17. An overview of the duration of the sequences is shown in Table 1. Figure 1 shows the same 3D sequence reconstructed using CS and CS-AI at the respective acceleration levels as well as the standard sense sequence as a reference. Upon review of the acquired images, it was discovered that one participant had a meniscal tear. Figure 2 shows the pathology on a 3D sequence with an acceleration factor of 10 reconstructed using CS and CS-AI compared to the reference sequence.

Fig. 1
figure 1

Comparison of a fat-saturated three-dimensional proton density sequence reconstructed using conventional compressed sensing (CS) and compressed sense with artificial intelligence (CS-AI) with acceleration levels of 10, 13, 15, and 17 as well as the standard sequence as reference (SENSE sequence with an acceleration factor of 2, included twice for easier comparison). Note the decreasing image quality with increasing acceleration factor. Overall, a better delineation can be observed in the lower row (CS-AI) compared to the upper row (CS)

Fig. 2
figure 2

Example of a meniscal tear: linear hyperintensity in the posterior horn of the meniscus. Almost no difference can be observed between the delineation of the lesion in the reference compressed sensing (CS) SENSE sequence (standard) compared to the CS with acceleration levels of 10 (CS-10) and CS-10 with artificial intelligence (CS-AI 10)

Subjective image analysis

Interrater agreement was assessed using Krippendorff´s α, indicating substantial interrater agreement for the subjective scoring for overall image impression at all acceleration factors (Krippendorff´s α = 0.71). The restricted maximum likelihood method-REML demonstrated a significant effect of the acceleration factors on the subjective measures of image quality (p < 0.001). Images reconstructed using CS-AI were rated significantly better than the respective sequences reconstructed using CS for all acceleration levels and all evaluated criteria (all p ≤ 0.001; see Table 3 and Fig. 3). Regarding the comparison to the reference sequence (CS sequence with an acceleration factor of 2), ratings for the sequences reconstructed using CS-AI did not differ significantly from the reference sequence for all evaluated criteria for an acceleration factor up to 10 (all p ≥ 0.999; see Table 3 and Fig. 3).

Table 3 Mean values and standard deviation for the subjective reading
Fig. 3
figure 3

Mean subjective ratings of the overall image quality for sequences reconstructed using compressed sensing (CS) and compressed sensing combined with a deep learning-based algorithm (CS-AI). The comparison between CS and CS-AI is shown in a, whereas b shows the comparison between CS-AI and the reference sequence (SENSE sequence with an acceleration factor of 2). * p ≤ 0.024, ** p < 0.010, *** p < 0.001

Discussion

The aim of the current study was to assess the subjective image quality of conventional CS and CS with AI at different acceleration factors for 3D diagnostic MRI of the knee. The purpose was to investigate the feasibility of using CS with AI to further decrease acquisition times of knee MRI while maintaining diagnostic image quality.

The findings demonstrate that images reconstructed using CS-AI consistently achieved significantly higher ratings for subjective measurements of image quality across all acceleration levels, compared to the corresponding images reconstructed using CS. These results align with previous research employing the same algorithm. Notably, Fervers et al. observed significantly improved subjective image quality for 3D T2-weighted images of the lumbar spine when employing CS-AI, as opposed to CS-based reconstruction [20]. Moreover, studies focusing on MRI of the ankle and prostate have reported enhanced objective and subjective image quality for sequences reconstructed with CS-AI [21, 22]. While CS-AI yields images with superior subjective image quality in comparison to CS, it is important to consider that excessively high acceleration levels may compromise diagnostic quality. Thus, our study aimed to assess the images generated by CS-AI across various acceleration levels and compare them to the reference sequence currently employed in clinical practice (a sense sequence with an acceleration factor of 2). Our objective was to determine an optimal acceleration factor that produces images with subjective image quality that is not significantly different from the existing clinical standard.

Based on the subjective image quality, a 3D sequence reconstructed using CS-AI with a tenfold acceleration did not perform significantly worse than the same sequence reconstructed using sense with an acceleration factor of 2. It is important to note that the other acceleration factors (13, 15, 17) did demonstrate a significant decline in image quality compared to the standard acquisition sequence. This decline was evident to a degree where these higher acceleration levels are not currently suitable to replace the standard sequence in clinical practice. This finding underscores the trade-off between higher acceleration and image quality, which is a critical consideration for the practical application of DL-based algorithms in MRI.

Replacing the standard sensitivity encoding sequence − SENSE 2 with the CS-AI 10 sequence (384 s versus 178 s) would result in 54% faster image acquisition for 3D sequences. Faster image acquisition holds the potential to enhance operational efficiency within imaging centers while enhancing patient accessibility to imaging services. Moreover, reducing the duration of patient scans can effectively minimize motion artifacts, leading to improved image quality, and therefore enhanced diagnostic accuracy. Moreover, short acquisition times can increase patient comfort during MRI, possibly making it more accessible to a broader range of patients. For instance, patients suffering from claustrophobia may find MRI more tolerable with reduced scan durations.

Nevertheless, it is essential to acknowledge the limitations of our study.

First, our sample size was relatively small, consisting solely of healthy volunteers. While the results from the two readers indicate comparable delineation of anatomical structures to the reference sequences, future investigations should involve scanning patients with prevalent knee pathologies to verify the preservation of image quality for pathological findings. An ongoing concern regarding DL-based reconstruction algorithms is the potential loss of information, where abnormal anatomical findings may be substituted with normal anatomy learned from the training data. To counteract this, the multi-scale network used in this study includes the integration of a data consistency term per coil element comparing the reconstructed data with the originally acquired data to ensure consistency [17]. Studies assessing other anatomical regions, including patients with pathologies and using the same CS-AI algorithm as in our study found no evidence for loss of information. For instance, Bischof et al. found no difference in Prostate Imaging Reporting and Data System − PI-RADS scores between images reconstructed using CS-AI compared to images reconstructed using CS [22]. Nevertheless, future studies should include more participants as well as patients with different pathologies to further evaluate the accuracy of algorithm.

Second, our study only focused on a 3D proton density sequence of the knee. Whereas other studies have shown similar performance of DL-based reconstruction algorithms across different MRI sequences [23], there are also studies showing that performance can differ between MRI sequences [24]. Thus, future studies should also include other sequences besides proton density sequences (e.g., T1-weighted sequences) to ensure that image quality are equally well preserved.

Third, it is important to note that our study did not incorporate objective measures of image quality. Previous research by Foreman et al. has discussed that traditional objective quality measures, such as signal-to-noise ratio, may not accurately reflect the quality of accelerated images generated through DL techniques [21]. Interestingly, Foreman et al. [21] observed an increase in signal-to-noise ratio with higher acceleration levels, despite a decrease in subjective image quality. This discrepancy could be attributed to the inherent noise reduction capabilities of DL-based reconstructions, rendering them less noisy compared to conventional reconstructions. Therefore, in our study, we relied solely on subjective image quality as the determining criterion to establish an appropriate acceleration factor that achieves comparable image quality to the reference sequence. Future investigations should explore alternative objective image quality measures that are not influenced by smoothing or denoising effects introduced by DL algorithms.

In sum, the results of our study show that the combination of DL and CS hold the potential for further scan time reduction in 3D imaging of the knee while providing overall better subjective image quality compared to CS alone. The implementation of this algorithm can help increase patient access to imaging and reduce motion artifacts by decreasing the overall time patients spend in the scanner. The results encourage further clinical investigation, extending the use cases to a clinical population and a wider range of MRI sequences. However, this study is limited by its small sample size and inclusion of only healthy volunteers, indicating the need for further research with a more diverse and larger sample.

Availability of data and materials

The data underlying this article cannot be shared publicly for the privacy of individuals that participated in the study. The data will be shared on reasonable request to the corresponding author.

Abbreviations

3D:

Three-dimensional

AI:

Artificial intelligence

CS:

Compressed sensing

DL:

Deep learning

MRI:

Magnetic resonance imaging

References

  1. Urwin M, Symmons D, Allison T et al (1998) Estimating the burden of musculoskeletal disorders in the community: the comparative prevalence of symptoms at different anatomical sites, and the relation to social deprivation. Ann Rheum Dis 57:649–655. https://doi.org/10.1136/ard.57.11.649

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  2. Baker P (2003) Knee disorders in the general population and their relation to occupation. Occup Environ Med 60:794–797. https://doi.org/10.1136/oem.60.10.794

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  3. Yawn BP, Amadio P, Harmsen WS et al (2000) Isolated acute knee Injuries in the general population. J Trauma 48:716–723. https://doi.org/10.1097/00005373-200004000-00021

    Article  CAS  Google Scholar 

  4. Heir S, Nerhus TK, Røtterud JH et al (2010) Focal cartilage defects in the knee impair quality of life as much as severe osteoarthritis. Am J Sports Med 38:231–237. https://doi.org/10.1177/0363546509352157

    Article  PubMed  Google Scholar 

  5. Gianotti SM, Marshall SW, Hume PA, Bunt L (2009) Incidence of anterior cruciate ligament injury and other knee ligament injuries: a national population-based study. J Sci Med Sport 12:622–627. https://doi.org/10.1016/j.jsams.2008.07.005

    Article  PubMed  Google Scholar 

  6. Oei EHG, Ginai AZ, Hunink MGM (2007) MRI for traumatic knee injury: a review. Semin Ultrasound CT MR 28:141–157. https://doi.org/10.1053/j.sult.2007.01.003

    Article  Google Scholar 

  7. OECD (2021) Magnetic resonance imaging (MRI) exams (indicator)

    Google Scholar 

  8. Subhas N, Benedick A, Obuchowski NA et al (2017) Comparison of a fast 5-minute shoulder MRI protocol with a standard shoulder MRI protocol: a multiinstitutional multireader study. AJR Am J Roentgenol 208:W146–W154. https://doi.org/10.2214/AJR.16.17041

  9. Griswold MA, Jakob PM, Heidemann RM et al (2002) Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med 47:1202–1210. https://doi.org/10.1002/mrm.10171

    Article  PubMed  Google Scholar 

  10. Lustig M, Donoho D, Pauly JM (2007) Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn Reson Med 58:1182–1195. https://doi.org/10.1002/mrm.21391

    Article  PubMed  Google Scholar 

  11. Bratke G, Rau R, Weiss K, et al (2019) Accelerated MRI of the Lumbar spine using compressed sensing: quality and efficiency. J Magn Reson Imaginghttps://doi.org/10.1002/jmri.26526

  12. Iuga AI, Abdullayev N, Weiss K, et al (2020) Accelerated MRI of the knee. Quality and efficiency of compressed sensing. Eur J Radiol 132:109273. https://doi.org/10.1016/j.ejrad.2020.109273

  13. Kijowski R, Rosas H, Samsonov A, et al (2017) Knee imaging: rapid three-dimensional fast spin-echo using compressed sensing. J Magn Reson Imaginghttps://doi.org/10.1002/jmri.25507

  14. Lin Z, Zhang X, Guo L et al (2019) Clinical feasibility study of 3D intracranial magnetic resonance angiography using compressed sensing. J Magn Reson Imaging 50:1843–1851. https://doi.org/10.1002/jmri.26752

    Article  PubMed  Google Scholar 

  15. Pezzotti N, Yousefi S, Elmahdy MS et al (2020) An adaptive intelligence algorithm for undersampled knee MRI reconstruction. IEEE Access 8:204825–204838. https://doi.org/10.1109/ACCESS.2020.3034287

    Article  Google Scholar 

  16. Iuga A-I, Rauen PS, Siedek F et al (2023) A deep learning-based reconstruction approach for accelerated magnetic resonance image of the knee with compressed sense: evaluation in healthy volunteers. Br J Radiol 96:20220074. https://doi.org/10.1259/bjr.20220074

    Article  PubMed  PubMed Central  Google Scholar 

  17. Pezzotti N, de Weerdt E, Yousefi S, et al (2019) Adaptive-CS-Net: FastMRI with adaptive intelligence. ArXiv. http://arxiv.org/abs/1912.12259

  18. Krippendorff K, Hayes AF (2007) Answering the call for a standard reliability measure for coding data. Commun Methods Meas 1:77–89. https://doi.org/10.1080/19312450709336664

    Article  Google Scholar 

  19. D’Agostino RB, Belanger A, D’Agostino RB (1990) A suggestion for using powerful and informative tests of normality. Am Stat 44:316–321. https://doi.org/10.1080/00031305.1990.10475751

    Article  Google Scholar 

  20. Fervers P, Zaeske C, Rauen P et al (2023) Conventional and deep-learning-based image reconstructions of undersampled k-space data of the lumbar spine using compressed sensing in MRI: a comparative study on 20 subjects. Diagnostics 13:418. https://doi.org/10.3390/diagnostics13030418

    Article  PubMed  PubMed Central  Google Scholar 

  21. Foreman SC, Neumann J, Han J et al (2022) Deep learning–based acceleration of compressed sense MR imaging of the ankle. Eur Radiol 32:8376–8385. https://doi.org/10.1007/s00330-022-08919-9

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  22. Bischoff LM, Katemann C, Isaak A et al (2023) T2 turbo spin echo with compressed sensing and propeller acquisition (sampling k-space by utilizing rotating blades) for fast and motion robust prostate MRI. Invest Radiol 58:209–215. https://doi.org/10.1097/RLI.0000000000000923

    Article  CAS  PubMed  Google Scholar 

  23. Lønning K, Putzky P, Sonke JJ et al (2019) Recurrent inference machines for reconstructing heterogeneous MRI data. Med Image Anal 53:64–78. https://doi.org/10.1016/j.media.2019.01.005

    Article  PubMed  Google Scholar 

  24. Eo T, Jun Y, Kim T et al (2018) KIKI-net: cross-domain convolutional neural networks for reconstructing undersampled magnetic resonance images. Magn Reson Med 80:2188–2201. https://doi.org/10.1002/mrm.27201

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

We acknowledge support for the Article Processing Charge from the DFG (German Research Foundation, 491454339) and the DEAL-Project.

Funding

Open Access funding enabled and organized by Projekt DEAL. The authors declare that this work has not received any funding.

Author information

Authors and Affiliations

Authors

Contributions

TD: conceptualization, investigation, formal analysis, writing — original draft, writing — review and editing, methodology, validation, data curation, visualization. CZ: conceptualization, investigation, formal analysis, writing — original draft, writing — review and editing, data curation. FS: conceptualization, investigation, formal analysis, writing — original draft, writing — review and editing, data curation. PR: conceptualization, investigation, formal analysis, writing — original draft, writing — review and editing, software. NGH: conceptualization, investigation, formal analysis, writing — original draft, writing — review and editing, data curation. KS: conceptualization, investigation, formal analysis, writing — original draft, writing — review and editing, visualization. DM: conceptualization, investigation, formal analysis, writing — original draft, writing — review and editing, visualization. GB: conceptualization, investigation, formal analysis, writing — original draft, writing — review and editing, supervision, methodology, validation. AI: conceptualization, investigation, formal analysis, writing — original draft, writing — review and editing, supervision, methodology, validation, data curation.

Corresponding author

Correspondence to Thomas Dratsch.

Ethics declarations

Ethics approval and consent to participate

This single center study was approved by our institutional review board and registered in the national Clinical Trials Register (DRKS00024156).

Consent for publication

Not applicable.

Competing interests

The author KS currently works for Philips GmbH Market DACH, Hamburg, Germany. DM and GB are on the speaker’s bureau of Philips Healthcare. The remaining authors have no potential conflict of interest to declare.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Dratsch, T., Zäske, C., Siedek, F. et al. Reconstruction of 3D knee MRI using deep learning and compressed sensing: a validation study on healthy volunteers. Eur Radiol Exp 8, 47 (2024). https://doi.org/10.1186/s41747-024-00446-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s41747-024-00446-0

Keywords